3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-17
Pulse Response
Adjustment
Equipment: Oscilloscope (2465B), 20dB Attenuator
1.
Configure the Model 3152A as follows:
Function:
Square
Frequency:
1 MHz
Amplitude:
10 V
Remote Commands:
:RES
:OUTP ON
:FUNC:SHAP
SQU
:FREQ
1e6
:VOLT
10
2. Connect the Model 3152A output through 20dB
attenuator to the oscilloscope input.
3.
Set the oscilloscope input impedance to 50
Ω
.
4.
Set the oscilloscope vertical gain and time base, and
adjust R36 and C19 for the best pulse response.
5.
Change the Model 3152A frequency setting to 10.00
KHz.
6.
Set the oscilloscope vertical gain and time base, and
adjust R22 for best pulse flatness.
7.
Change the Model 3152A frequency setting to 1.000
KHz and retouch R22 for best flatness in this range.
Repeat Steps 5 and 6 until the best result is obtained in
both steps.
Amplitude
Adjustment
Equipment: DMM, 50
Ω
feedthrough termination
1.
Configure the Model 3152A as follows:
Function:
Sine
Frequency:
1 KHz
Amplitude:
16 V
Remote Commands:
:RES