3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-19
both steps.
VCO Adjustment
Equipment: DMM
1.
Configure the Model 3152A as follows:
Function:
Sinewave
Output:
On
Remote Commands:
:RES
2.
Connect the DMM “V” input to U515 pin 1 and the DMM
“common” nput to case ground. Set the DMM to DCV
measurements.
3.
Adjust R536 until the DMM reading is 0 V
±
500 mV.
PLL Adjustment
Equipment: Oscilloscope, Counter, Pulse/Function Generator,
Feedthrough Terminator, 3 each BNC Cables, 50
Ω
Tee
Connector
1.
Configure the Model 3152A as follows:
Function:
Square
wave
Mode:
Front Panel PLL
Frequency:
10.00
KHz
Mode:
PLL
Phase Offset
45
°
Amplitude:
2
V
Trigger Level:
0 V
Remote Commands:
:RES
:OUTP ON
:FUNC:SHAP
SQU
:FREQ 10.00E3
:PHAS2:LOCK ON
:PHAS2:ADJ 45
:TRIG:LEV 0