3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-13
PLL
Characteristics
Accuracy: 5.4 x frequency (in MHz) + resolution / 2
±
5
°
Equipment: Oscilloscope, Counter, Pulse/Function Generator,
Feedthrough Terminator, 3 BNC Cables, 50
Ω
Tee Connector
1.
Configure the Model 3152A as follows:
Function:
Square wave
Mode:
Front Panel PLL
Frequency:
1.000 KHz
Amplitude:
2 V
Trigger Level:
0 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP
SQU
:PHAS2:LOCK
ON
:TRIG:LEV 0
:VOLT
2
2.
Make the following connections:
a. Connect function generator output to the input of
the 50
Ω
Tee Connector.
b. Connect one end of the 50
Ω
Tee Connector to the
3152A TRIG IN and the other end to the counter
Channel A input.
c. Connect 3152A OUT to the counter Channel B
input.
3.
Set counter to phase measurements. Terminate counter
inputs with 50
Ω
.
4.
Set Pulse/Function generator to output 2 Vp-p square
waveforms. Change 3152A and Pulse/Function generator
frequency and verify counter phase readings as shown
below.