There are two ways to load waveforms into the Model 3152A;
using a graphical user interface, i.e., WaveCAD, or low-level
programming. When using WaveCAD, disregard most of this
chapter as WaveCAD does the work for you. When writing your
own code, use the following commands to load data into a
specific memory segment.
First, define the work area. Define the segment number and its
associated length. Segment length must be an even number.
For example, to use segment number 8 and give it a length of
1000 points, use this command:
TRACe:DEFine
8,1000
Note
When large numbers of traces need to be defined it
may be beneficial to use the SEGM:DATA command
in place of repeated calls to TRAC:DEF when
throughput needs to be optimized (see the next
section for details).
Next, make segment 8 the active segment. The active segment
must be selected because as waveforms are loaded, the Model
3152A must be notified as to where to place the data it receives.
Select the active segment using the following command:
TRACe: SELect 8
The next step is to transfer data to the active segment. Data is
loaded into the Model 3152A using high-speed binary transfer. A
special command is defined by IEEE-STD-488.2 for this
purpose. High speed binary transfer allows any 8-bit bytes
(including extended ASCII code) to be transmitted in a message.
This command is particularly useful for sending large quantities
of data. The Model 3152A uses this command to receive
waveforms from the controller:
TRACe
#42000<binary_block>
This command causes the transfer of 2000 bytes of data (1000)
points into the active memory segment. The ASCII "#" ($23) is
the start of the binary data block. "4" designates the number of
digits that follow. "2000" is the even number of bytes to follow.
The generator represents binary data as 12-bit integers that are
sent as two bytes. Therefore, the total number of bytes is always
twice the number of data points in the waveform. For example,
2000 bytes are required to download a waveform with 1000
points. Bytes are sent in byte-high, byte-low order. The
FORM:SWAP command can be used to reverse this order.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...