3152A User Manual
Racal Instruments
©
2001
Using The Instrument 3-19
The selected standard waveform will be available at the output
connector only after the FUNC:MODE FIX command has been
executed.
With the FUNC:SHAP PULS selection, parameters for the
PULSe function will effect the output waveform when modified.
Note that changes made to parameters for a specific function do
not have any effect on other functions. Also, changes made to
other function parameters while the pulse shape is selected will
not have any effect until that pulse shape is selected.
The command:
PULSe: DELay <value>
Sets the pulse delay in percent of the pulse period. For example,
if the pulse period is 100 ms, 10% will delay the first transition of
the pulse by 10 ms. Delay is measured from trigger to the first
turning point.
The query:
PULSe:DELay?
queries the pulse delay setting and returns a value in percent.
The command:
PULSe: TRANsition <value>
sets the pulse rise time in percent of the pulse period. For
example, if the pulse period is 100 ms, 5% rise time equals 5
ms. Pulse rise time is measured between the two turning points
of the first transition.
The query:
PULSe:TRANsition?
queries the pulse rise time setting and returns a value in percent.
The command:
PULSe: WIDTh <value>
sets the pulse width in percent of the pulse period. For example,
if the pulse period is 100 ms, 20% pulse width equals 20 ms.
Pulse width is measured between the two turning points on the
top of the pulse. The query:
PULSe:WIDTh?
queries the pulse width setting and returns a value in percent.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...