3152A User Manual
Maintenance And Performance Checks 5-18 Racal Instruments
©
2001
:OUTP ON
:FREQ 1E3
:VOLT
16
2.
Connect the Model 3152A output to the DMM input
through the 50
Ω
feedthrough termination. Set the DMM
to AC Volt measurements and 2 V range.
3.
Adjust R44 until the DMM reading is 5.657 V
±
20 mV.
4.
Change the Model 3152A amplitude setting to 2.0 V.
5.
Adjust R44 until the DMM reading is 0.707V
±
6 mV.
6.
Repeat Steps 4 and 5 until the best result is obtained in
both steps.
Offset Adjustment
Equipment: DMM, 50
Ω
feedthrough termination
1.
Configure the Model 3152A as follows:
Function:
Sine
Frequency:
1 MHz
Amplitude:
1.61 V
Offset:
+7.19V
Remote Commands:
:RES
:OUTP ON
:FREQ
1e6
:VOLT
1.61
:VOLT:OFFS
7.19
2.
Connect the Model 3152A output to the DMM input
through the 50
Ω
feedthrough termination. Set the DMM
to DCV measurements and 10 V range. Record this
reading.
3.
Change the offset setting to -7.19 V and adjust R85 to
have the exact reading as was recorded in the previous
step (with reversed polarity).
4.
Adjust R62 until the DMM reading is -7.190 V
±
30 mV.
5.
Repeat Steps 3 and 4 until the best result is obtained in
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...