Racal Instruments
©
2001
Maintenance and Performance Checks 5-11
2.
Connect the Model 3152A OUTPUT to the oscilloscope,
set the oscilloscope input impedance to 50
Ω
, and set
the oscilloscope to display the sinewave within exactly 6
vertical divisions.
3.
Change the Model 3152A FREQ setting to 1.000 MHz.
Verify that peak-to-peak of the displayed sinewave is
within 5.9 to 6.1 divisions.
4.
Change the Model 3152A FREQ setting to 10.00 MHz.
Verify that peak-to-peak of the displayed sinewave is
within 5.7 to 6.3 divisions.
5.
Change the Model 3152A FREQ setting to 50.00 MHz.
Verify that peak-to-peak of the displayed sinewave is
within 5.1 to 6.9 divisions.
Trig, Gate and
Burst
Characteristics
Triggered
-
Each transition at the front panel TRIG IN connector
stimulates the Model 3152A to generate one complete output
waveform.
Gated
-
External signal at the TRIG IN connector enables the
Model 3152A output. The last cycle of the output waveform is
always completed.
Burst
-
Each transition at the front panel TRIG IN connector
stimulates the Model 3152A to generate a burst of pre-selected
number of cycles.
Equipment
:
Pulse/function generator (8021), oscilloscope.
Trigger
1.
Configure the Model 3152A as follows:
Function:
Sine wave
Frequency:
1.000 MHz
Operating Mode:
Triggered
Remote Commands: :RES
:OUTP ON
:FREQ
1e6
:INIT:CONT
OFF
2. Set the external pulse/function generator frequency to
10 KHz and connect its SYNC connector. Set the
oscilloscope and verify on the oscilloscope that the
Model 3152A outputs a triggered signal. Leave the
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...