3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-15
Adjustments
Environmental
Conditions
Adjustments should be performed under laboratory conditions
having an ambient temperature of 25
o
C
±
5
o
C and relative
humidity of less than 80%. If the instrument has been outside
this range, allow at least one additional hour for the instrument
to stabilize before beginning the adjustment procedure.
Between adjustments, always leave the top cover on the unit to
keep the internal temperature as stable as possible.
Warm-Up Period
Most equipment is subject to a small amount of drift when first
turned on. To ensure long term calibration accuracy, turn on the
power to the Model 3152A and allow it to warm-up for at least
30 minutes before beginning the adjustment procedure.
Recommended
Test Equipment
Recommended test equipment for calibration is listed at the
beginning of this chapter in the section entitled
Recommended
Test
Equipment
. Test instruments other than those listed may
be used only if their specifications equal or exceed the required
characteristics.