Maintenance and Performance Checks 5-5
Recommended
Test Equipment
Recommended test equipment for troubleshooting, calibration
and performance checking is listed below. Test instruments
other than those listed may be used only if their specifications
equal or exceed the required characteristics.
Test Instrument & Accessories
Minimum Specification
Universal
Counter
100MHz,
10ppm
accuracy
DMM
ACV,
0.2%;
DCV,
0.1%
accuracy
Distortion
Analyzer
10
Hz
to
100KHz
Spectrum
Analyzer
1GHz
bandwidth,
1KHz
resolution
Pulse/Function
Generator
20
MHz,
10ns
pulsewidth
Oscilloscope
2
channels,
400MHz
analog
bandwidth
Feedthrough
termination
50
Ω
, 1%
BNC
Cables
1
meter
long,
50
Ω
impedance
20dB
feedthrough
attenuator
50
Ω
, 1%
“T” BNC Connector
50
Ω
, 1%
Performance
Check Procedures
Use the following procedures to check the Model 3152A against
the specifications. A complete set of specifications is listed in
Appendix A. The following paragraphs show how to set up the
instrument for the test, what the specifications for the tested
function are, and what acceptable limits for the test are. If the
instrument fails to perform within the specified limits, the
instrument has to be calibrated or tested to find the source of
the problem.
Frequency
Accuracy
Accuracy Specifications
:
0.01% of reading to 50.00 MHz
Equipment
:
Counter
1.
Configure the Model 3152A as follows:
Function:
Square
Frequency:
10 Hz
Amplitude:
1 V
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...