3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-23
Table 5-1, CPU and VXI Interface Checkout Procedure
Step
Test Point
Expected Result
1
CPU Clock - U87 pin 19
16 MHz TTL level clock
2
VXI Clock - U109 pin 159
16 MHz TTL level clock
3
ALE - U85 pin 11
Positive going pulses; period 375 ns,
pulse width 120 ns
4
PSEN - U74a and U74b
2.67 MHz TTL level pulses
5
Load a 10000-point wave and
perform the following tests during
load time
6
RD - U87 pin 16
Negative going 400 ns width pulses
7
WR - U87 pin 17
Negative going 400 ns width pulses
8
LDACK - U87 pin 5
Negative going 6.5
μ
s width pulses
9
ASIC LCS - U104 pin 11
Negative going 6.5
μ
s width pulses
10
LR/W - U93 pin 19
Negative going 60
μ
s width pulses
Output Amplifier
and Amplitude
Control Checkout
Problems with the output amplifier and amplitude control circuits
can be detected if the Model 3152A accepts word serial
commands and responds correctly to word serial queries, but
fails to output the correct waveform with correct amplitude and
offset levels. If there is a problem with the output amplifier or
with the amplitude control circuit, proceed with the checkout
procedure given in
Table 5-2.
Before starting the following procedure, reset the Model 3152A
to the factory defaults using the *RST command. Then send
the command OUTPUT ON.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...