Using The Instrument 3-45
Figure 3-10, Sequenced Waveforms - Triggered Advance Mode
Triggered
Sequence
Advance
Commands
Placing the Model 3152A in Triggered Sequence Advance
mode is done in Triggered mode only. First, prepare the
sequence of waveforms using the commands that were
explained before. Second, place the instrument in Triggered
mode using the INIT:CONT OFF command. To place the Model
3152A in Triggered Sequence Advance mode, use the following
command:
SEQuence:ADVance {AUTO | TRIGger}
AUTO specifies the normal continuous advance and TRIGger
places the instrument in Triggered Sequence Advance mode.
The query:
SEQuence:ADVance?
queries Triggered Sequence Advance mode and returns AUTO
or TRIG.
Backplane Inter-
Module
Synchronization
Although multiple Model 3152As within one chassis run off a
common clock (CLK10), their outputs are not synchronized to
each other. If the same waveform length and clock rates for
two modules are selected and both are displayed on an
oscilloscope, the outputs may look as though they are
synchronized even though they are not. The waveforms may
not start at the same point along the waveform. If another
waveform is selected or if the same waveforms are
reprogrammed, you may notice that the phase relationship
between the two modules has changed again.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...