3152A User Manual
Racal Instruments
©
2001
Using The Instrument 3-9
The query:
FUNCtion:MODE?
queries the output function type and returns either FIX, USER,
SEQ or SWE.
Selecting a
Standard
Function Shape
Use the following command to select a standard output function:
FUNCtion: SHAPe {SINusoid | TRIangle | SQUare |
PULSe | RAMP | SINC | EXPonential | GAUSsian | DC}
The selected waveform is output using the previously selected
frequency, amplitude, offset, and other relevant settings. The
standard waveform will be output only after the
FUNC:MODE:FIX command is selected.
The query:
FUNCtion:SHAPe?
queries the standard function shape and returns either SIN, TRI,
SQU, PULS, RAMP, SINC, EXP, GAUS or DC.
Changing the
Frequency and
Sample Clock
Use the following command to change the frequency for
standard waveforms and sample clock for arbitrary waveforms:
FREQuency {<frequency> | MINimum | MAXimum}
MIN selects the lowest frequency allowed for the currently active
function. MAX selects the highest frequency allowed for the
currently active function. The default frequency setting is 1 MHz
for all functions.
The query:
FREQuency?
queries the frequency setting for the standard function currently
active and returns a value in Hertz.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...