3152A User Manual
Using The Instrument 3-14
Racal Instruments
©
2001
Enabling the Main
Output
For safety reasons, the main output default setting is OFF.
Disable or enable the main output using the following command:
OUTPut {OFF | ON}
When the main output state is programmed to ON, the output
connector is connected to the output amplifier through a 50
Ω
resistor. In the OFF position, the output connector is
disconnected from the output amplifier by means of a
mechanical relay. Ensure that voltage is
not
applied to the main
output connector when the Model 3152A output state is
programmed to ON.
The query:
OUTPut?
queries the state of the main output and returns "0" (OFF) or "1"
(ON).
Enabling the
SYNC Output
For safety reasons, the SYNC output default setting is OFF.
Disable or enable the SYNC output using the following
command:
OUTPut: SYNC {OFF | ON}
When the SYNC output state is programmed to ON, the SYNC
output connector generates signals which are triggered by
signals selected using the SYNC:SOUR command. In the OFF
position, the SYNC connector has no output. It is connected
electrically to the internal circuitry at all times. Ensure that
voltage is
not
applied to the SYNC at any time. The default
SYNC position is the 6th point from the end of the waveform.
The position of the SYNC signal can be programmed using the
OUTPUT:SYNC:POS:POIN command.
The query:
OUTPut:SYNC?
queries the state of the SYNC OUTPUT and returns "0" (OFF) or
"1" (ON).
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...