3152A User Manual
Using The Instrument 3-10
Racal Instruments
©
2001
The
command:
FREQuency: RASTer {<frequency> | MINimum |
MAXimum}
sets the sample clock frequency for the user and sequenced
functions. MIN selects the lowest frequency allowed for the
currently active segment or sequence. MAX selects the highest
frequency allowed for the currently active segment or sequence.
The default sample clock frequency setting is 1 MHz for all
functions.
Note that the output frequency depends on the number of points
specified in the waveform. The output frequency can be
computed using the following formula: Output Frequency =
Sample Clock / Number of points in the active segment.
The query:
FREQuency:RASTer?
queries the sample clock frequency setting for the arbitrary
segment or sequence currently active and returns a value in
Hertz.
Selecting the
Sample Clock
Source
Use the following command to select the source for the sample
clock for the user and sequenced functions:
FREQuency:RASTer:SOURce {EXT | INT | ECLTrg0}
EXT selects an external clock source. The external source is
applied to the front panel CLOCK IN connector. INT selects the
internally synthesized clock synthesizer. ECLTrg0 selects a
sample clock that is available on the backplane. Note that
ECLTrg0 is always the active sample clock source when the
Model 3152A is set to operate as a slave in phase
synchronization mode.
The query:
FREQuency:RASTer:SOURce?
queries the sample clock source setting and returns EXT, INT or
ECLT0.
Содержание 3152A
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 18: ...3152A User Manual xii This page was left intentionally blank...
Страница 25: ...3152A User Manual Racal Instruments 2001 Getting Started 1 7 Figure 1 2 3152A Front Panel...
Страница 156: ...3152A User Manual Product Support 6 4 Racal Instruments 2001 This page was left intentionally blank...