52
Preparing Your Laboratory
UPS Systems
The PowerVar 10.0 KVA On-Line Conditioned UPS 50/60Hz unit is currently
available for the THGA PinAAcle models (900T/Z). It meets the minimum surge
current handling rating of 90 A for about 3 seconds during Atomization stage peak
power. The PerkinElmer part number for the 10.0 KVA UPS is N0777613.
The PowerVar 800 VA On-Line Conditioned UPS 50/60Hz unit is currently
available for the PinAAcle 900F Flame Only model. The PerkinElmer part number
for the 800 VA UPS is N0777681.
5.2 KVA Line Conditioned UPS 50/60Hz unit is currently available for the HGA
PinAAcle model 900H. The PerkinElmer part number for the 5.2 KVA UPS is
N0777511.
Power Outlets
In addition to the above, the following are required:
•
For connection of the Cooling System, a 230 VAC socket is provided at
the rear of the PinAAcle Series 900 T/H/Z models.
•
For the PC, monitor, printer, and addition accessories, e.g., FIAS or Auto
Prep 50, suitable outlets must be provided. Protecting earth and ground
must be the same as for the PinAAcle Series 900 T/H/Z models (avoid
ground loops).
Gas Requirements
Note
Standards for cylinder dimensions, regulator connections, gas fittings, etc.
vary from country to country. The information provided here is for the
U.S. Contact your PerkinElmer representative for information on the
specific requirements of your area.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......