23
Safe Handling of Gas Cylinders
Identification of Gas Cylinders
•
Legibly mark cylinders to identify their contents. Use the chemical name or
commercially accepted name for the gas.
Storing Cylinders
Review the following precautions with the customer to ensure the safe use and
storage of gas cylinders.
•
Cylinders should be stored in accordance with the regulations and standards
applicable to the customers locality, state, and country.
•
Flammable gases must be stored separately from the non-flammable gases and
oxidizing gases stored separately from the non-oxidizing gases.
•
When cylinders are stored indoors in storage rooms, the storage room should
be well ventilated and dry. Ensure that the ventilation is adequate to prevent
the formation of dangerous accumulations of gas. This is particularly important
in small or confined areas.
•
Do
not
store cylinders near elevators, gangways, or in locations where heavy
moving objects may strike or fall against them.
•
Use and store cylinders away from exits and exit routes.
•
Locate cylinders away from heat sources, including heat lamps. Compressed
gas cylinders should not be subjected to temperatures above 52 °C (126 °F).
•
Do
not
allow ignition sources in the storage area and keep cylinders away from
readily ignitable substances such as gasoline or waste, or combustibles in bulk,
including oil.
•
Store cylinders standing upright, fastened securely to an immovable bulkhead
or permanent wall.
•
When storing cylinders outdoors, they should be stored above ground on a
suitable floor and protected against temperature extremes (including the direct
rays of the sun).
•
Mark gas cylinders clearly to identify the contents and status (full, empty, etc.).
•
Do
not
attempt to refill gas cylinders.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......