250
Maintenance
Lamp Maintenance and Cleaning
There are really no maintenance procedures for any of the PerkinElmer lamps. If
you notice fingerprints on the "window" of your lamp, wipe them off with a soft,
dry, lint free cloth. Do not use chemicals or water. See the section below regarding
lamp life and when to replace them.
Lamp Life
Both hollow cathode and electrodeless discharge lamps have a finite lifetime.
Electrodeless discharge lamps typically last longer than hollow cathode lamps for
the same element. If, while aligning a lamp, you are not able to obtain a satisfactory
energy value, this may indicate that the lamp is starting to deteriorate. As a hollow
cathode lamp reaches the end of its useful life, the current on the lamp may be
increased up to the maximum current recommended on the label. Once the
intensity of a lamp begins to deteriorate, it should be replaced. The Lumina lamps
comes with an electronic means of monitoring its usage, also in mA-Hours, with
an in-built EEPROM, that can be addressed and read from the PinAAcle AA.
PerkinElmer lamps are warrantied for 5000 mA-hour or six months whichever
comes earlier.
Lamp Troubleshooting
Refer to Chapter 8, Troubleshooting, in this manual for lamp troubleshooting
information.
Replacing the Lamps
Caution
To avoid damages to the instrument optics,
do not
insert uncoded non-
Lumina lamps, such as Intensitron lamps, too far into the lamp holder.
Always wear UV-absorbing safety glasses when installing, replacing or
aligning hollow cathode lamps or electodeless discharge lamps.
The ultraviolet radiation emitted by these lamps can harm your eyes.
!
Warning
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......