244
Maintenance
Cleaning the Nebulizer
1. After you have taken the nebulizer apart, clean the components as follows:
2. Clean the capillary assembly using a mild laboratory detergent solution, or a
solvent such as isopropanol, preferably in an ultrasonic bath.
3. Clean the nebulizer body and any other parts that appear dirty in a mild
laboratory detergent solution, or a solvent such as isopropanol, preferably in an
ultrasonic bath.
4. Rinse all the cleaned components thoroughly with deionized water.
5. Inspect the all o-rings and clean them if necessary in mild laboratory detergent
solution, followed by a thorough rinsing with deionized water. If they look
damaged, replace them.
6. After you have cleaned the components, reassemble the nebulizer.
Removing and Replacing the High Sensitivity Plastic Nebulizer's
Impact Bead
The PinAAcle plastic nebulizer utilizes a ceramic, one-piece, impact-bead
structure (Part No. B0505086) which improves nebulization efficiency. The bead
itself is held by two fragile support arms.
The bead structure mounts directly on the plastic nebulizer's body assembly, see
the figure on page 239.
To Remove the Impact Bead
1. Hold the impact bead by its base only.
2. Rotate it approximately 1/8 turn counter clockwise before carefully lifting it
from the nebulizer's body assembly.
To Install the Impact Bead
1. Hold the impact bead by its base and carefully push it onto the body of the
plastic nebulizer in front of the venturi.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......