143
Installing the Sample Capillary
Table 4-2.
Sample cups and reagent containers
•
Polypropylene containers are suitable for organic solvents
Installing the Sample Capillary
To avoid any damage to the sample capillary during shipment, the sample capillary
will have to be removed from the pipette arm assembly.
To install the sample capillary:
1. Carefully remove any tape used for holding the capillary during transportation.
2. Push the metal mount of the sample capillary into the clip at the end of the
pipette arm. Make sure that the sharp side of the metal mount goes first.
3. Pass the sample tube through the hooks on the pipette arm such that it forms a
smooth loop from the metal mount.
4. Pass the sample tube around the hooks on the sample table, leaving a
sufficiently large loop so that the pipette arm can move freely.
Filling and Flushing the Rinsing System
The quality of the rinsing and sample delivery system of the AS 900 autosampler
is most important to achieve good analytical results. Flushing is necessary after all
alterations to the system (e.g. new sample capillary). after changing the sample
Cup or container
88-location tray
148-location tray
volume, material
locations
locations
1.2 mL; polypropylene
not suitable
1–120
122, 123, 125, 127, 128,
130, 132, 133, 135, 137,
138, 140
2.5 mL; polypropylene
1–80
121, 124, 126, 129, 131,
134, 136, 139
7 mL; polypropylene
81–88
141–148
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......