349
Performance Problems
Rapid corrosion of tube
Argon flow rate is too low.
Make sure that the argon gas
(or Nitrogen gas, or any insert
gas) is clean and DRY. The
supply - piping and tubing, of
this clean and DRY gas to the
furnace must also be clean
and dry.
Set suitable flow rates in each
step of the furnace program.
Set the correct inlet pressure
for the argon.
The drying step is not ade-
quate. Solvent fumes remain
in and around the tube after
drying, when raising to higher
temperatures of pyrolysis and
atomization.
With WinLab32 for AA, use
the Method Editor to enter a
suitable furnace program. See
the Recommended Condi-
tions.
The samples/solutions ana-
lyzed may be very corrosive,
and/or also contain high con-
centration of acids, so the
tube life is shorter.
Dilute the solutions, to reduce
its corrosiveness, the acid
concentration, and to use
appropriate temperature pro-
gram that may contain slower
ramp, lower pyrolysis and
atomization temperatures.
Develop equitable tempera-
ture program. Please seek
advice from the PerkinElmer
Product Specialists.
Autosampler does not go to
the correct locations.
The wrong configuration of
the tray for the autosampler.
Select the correct sample tray
in
WinLab32 > Options >
Furnace Autosampler
. See
the online help.
Autosampler does not oper-
ate.
The communication cable is
not connected correctly.
Make sure that you use the
correct cable, and connect it
to the correct communication
port.
Problem
Possible Cause(s)
Corrective Action
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......