44
Preparing Your Laboratory
Exhaust venting is important for the following reasons:
•
It will protect laboratory personnel and the instrument from toxic vapors,
which may be produced by some samples.
•
It will minimize the effects of room drafts and the laboratory atmosphere
on flame stability.
•
It will help to protect the instrument from corrosive vapors which may
originate from the samples.
•
It will remove dissipated heat produced by the flame or furnace.
We strongly recommend that you do not place the spectrometer in a chemical
hood.
If a chemical hood must be used, make arrangements to prevent
contamination due to corrosive vapors and back drafts from other hoods.
Do
not
perform sample preparation in the same hood as the spectrometer.
Vent Hood Dimensions
For Vent Hood Alignment - Approximately 58 cm (23 in) from back of Instrument
to center of Atomizer. Approximately 29 cm (11.5 in) from tight side of Instrument
to center of Atomizer.
Warning: Toxic Gases - Fume Ventilation System
Without adequate ventilation, potentially toxic vapors can build up in the
laboratory.
Your laboratory must have a reliable fume ventilation system before
you use this instrument.
!
Warning
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......