213
Cleaning the Burner System
•
Aspirating samples with high solids content may cause deposits to build up
along the burner slot
•
Prolonged operation of the nitrous oxide-acetylene flame may cause deposits
of pyrolytic carbon to build up along the burner slot
Flashbacks
A flashback is an explosion of the premixed gases in the burner chamber. The
safety interlocks normally prevent conditions that could cause a flashback, but if a
flashback should ever occur:
•
Check that the burner head slot is clean
•
Check that the burner head is seated correctly
•
Check the flow spoiler or impact bead for damage, and replace any damaged
parts
•
Check the inside of the end cap for any cracks.
Unstable acetylides, which are likely to explode when dry, may be
formed when aspirating high concentrations of silver, copper, or mercury
salts into an acetylene flame.
Warning: Hydrochloric and Nitric Acids - Risk of Burns
Hydrochloric (HCl) and nitric (HNO
3
) acids are corrosive and can cause
severe burns.
Caution
If high concentrations of Cu, Ag, or Hg have been aspirated into an
acetylene flame,
clean the burner chamber immediately after use
by
following the procedures in this section called
Cleaning the Burner After
Aspirating Cu, Ag, or Hg sample
.
!
Warning
!
Warning
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......