53
Gas Requirements
Gas Line Connections
PerkinElmer AA instruments include the hoses necessary for connection to gas
lines. It is the responsibility of the user to provide the appropriate gas lines,
regulators, connectors and valves to which the hoses are connected.
Safety of Gas Cylinders and Gas Delivery Lines
While the installation and operation of all gas supplies is the responsibility of the
user, the following safety information concerning the handling of gas cylinders is
included for your reference:
•
Flammable gas cylinders (such as acetylene) should not be stored with
oxygen, or nitrous oxide cylinders, or adjacent to oxygen charging
facilities.
•
Fasten all gas cylinders securely to an immovable bulkhead or a permanent
wall.
•
When gas cylinders are stored in confined areas, such as a room,
ventilation should be adequate to prevent toxic or explosive
accumulations. Move or store gas cylinders only in a vertical position with
the valve cap in place.
•
Locate gas cylinders away from heat or ignition sources, including heat
lamps. Cylinders have a pressure-relief device that will release the
contents of the cylinder if the temperature exceeds 52 °C (125 °F).
•
When storing cylinders external to a building, the cylinders should be
stored so that they are protected against temperature extremes (including
the direct rays of the sun) and should be stored above ground on a suitable
floor.
•
Mark gas cylinders clearly to identify the contents and status (full, empty,
etc.).
•
Do not attempt to refill gas cylinders.
Note
The permanent installation of gas supplies is the responsibility of the user
and should conform to local safety and building codes.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......