360
Troubleshooting
Poor precision
Poor drying conditions caus-
ing sample spattering.
No rinse solution
Remedy is to carefully moni-
tor the drying with a dental
mirror and modify the time/
temp program in WinLab32
for AA as needed.
Fill the rinse bottle and prime
the rinsing system correctly.
The autosampler pipet tip is
partly blocked or damaged.
Make sure that the pipet tip
and the sample tube are clean
and undamaged. Replace
them if necessary. See
Furnace Autosampler Main-
tenance on page 322.
Air bubbles in the pumps
Flush the air out of the
pumps. Prime the rinsing sys-
tem correctly. See
Furnace Autosampler Main-
tenance on page 322
Valves: dirty or worn.
Clean the valves or fit new
valves. See the user's guide
for the furnace. See
The furnace is not correctly
aligned.
Align the furnace. See
Graphite Furnace Mainte-
nance for the PinAAcle 900T/
900Z on page 260
Erratic signals,
double peaks.
In some situations, multiple
peaks cannot be avoided. The
errors from multiple peaks are
much less pronounced when
integrated absorbance is used
as opposed to peak height.
In many situations, raising the
atomization temperature will
eliminate the multiple peaks.
Use appropriate matrix modi-
fiers, and Temperature Pro-
gram.
Problem
Possible Cause(s)
Corrective Action
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......