180
Preparing the System
To rinse the system:
The procedure for rinsing the burner system depends on the type of samples that
you have analyzed.
Recommendations:
•
If the sample solutions contained organic solvents, use a multi-rinse procedure
to remove all traces of the solvents from the system.
•
If the sample solutions contained toxic substances, you must rinse all traces of
these substances out of the system.
•
If the sample solutions contained high concentrations of copper, silver or
mercury salts, which can form unstable acetylides, you must rinse all traces of
these substances out of the system.
Procedure:
•
If you used only aqueous solutions during the analyses:
1. Aspirate deionized water for five minutes.
•
If you used organic solvents during the analyses:
2. Aspirate for five minutes an organic solvent that is miscible with the samples
that have just been aspirated.
3. Aspirate for five minutes methanol or another solvent that is miscible both with
water and the solvent used previously.
4. Aspirate 1% (v/v) nitric acid solution for five minutes.
5. Aspirate deionized water for five minutes.
To extinguish the flame:
1. With the flame still burning, aspirate the correct rinsing solutions to rinse the
burner system. Either aspirate the solutions manually or use the flame
autosampler.
2. In the Flame Control window, click on the
Flame
icon.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......