221
Reassembling the Burner System
3. Remove the nebulizer from the end cap assembly. Refer to
Stainless Steel Nebulizer on page 231
4. Clean the chamber. Place the parts in a detergent solution until they are
Cleaning the Burner Chamber on page 208
5. Remove and rinse with deionized water. Allow time for complete drying.
Reassembling the Burner System
After you have completed maintenance procedures, reassemble the burner system
as given below.
To reassemble the burner chamber
1. Before you reassemble the burner system, verify that the o-rings are in good
condition. If an o-ring appears damaged, replace it.
2. Connect the drain tube and clamp.
3. Carefully slide the nebulizer into the end cap and lock into place with the
nebulizer clamp.
4. Place the burner o-ring into the burner chamber and screw on the retaining ring.
5. Press the safety latch on the ignitor box while gently pushing and twisting the
burner head onto the chamber.
To Install the Burner in the Spectrometer
1. Guide the burner assembly into the tracks of the Burner Assembly mount.
2. Push the burner assembly in so that the rear gas fittings line up with the gas
connectors at the back of the Burner Assembly mount. Push the burner
assembly fully in until it clicks into place. Pull the locking level forward until
it locks the burner assembly in place. The locking pin automatically locks.
Содержание PINAACLE 900 Series
Страница 1: ...PINAACLE 900 SERIES Customer Hardware and Service Guide ATOMIC SPECTROSCOPY...
Страница 2: ......
Страница 3: ...PinAAcle 900 Series Customer Hardware and Service Guide...
Страница 26: ...10 Safety Practices Burner Locking Mechanism Label Item Description 1 Press button to unlock the lever 1...
Страница 52: ...36 Safety Practices...
Страница 55: ...39 Lab Space Requirements Dimensions and Weights Figure 2 1 Instrument dimenstions...
Страница 84: ...68 Preparing Your Laboratory...
Страница 91: ...75 The PinAAcle 900 Series Spectrometer System Figure 3 2 Optical System for the PinAAcle 900H...
Страница 92: ...76 System Description Figure 3 3 Optical System for the PinAAcle 900T...
Страница 93: ...77 The PinAAcle 900 Series Spectrometer System Figure 3 4 Optical System for the PinAAcle 900Z...
Страница 101: ...85 Burner Assembly System Figure 3 8 Block diagram of the burner system...
Страница 210: ...194 Preparing the System...
Страница 278: ...262 Maintenance Figure 6 31 Insertion removal tools Item Description 1 Graphite tube holder insertion removal tool 1...
Страница 310: ...294 Maintenance Figure 6 54 Window key Item Description 1 Window key 1...
Страница 354: ...338 Maintenance...
Страница 388: ...I 6 Index W WEEE Instructions 34 Weights 39 Windows cleaning 321 refitting 320 removing 318...
Страница 389: ......