Technical Data 23
Gate Position:
•
Start: 1.16 to 185" (-29 mm to 4694 mm) of steel (-10 to 1600 µs).
•
Width: 0.001 to 296" (0.001 mm to 7510 mm) of steel (0.001 to 2560 µs).
•
Control: Continuous over full range adjustment.
•
Display: Alarm bar indicating position and threshold.
•
Accuracy: ±1% of indicated position.
Threshold
•
Control: Variable, 5 to 100% of full screen.
•
Accuracy: ±1%.
Flaw Alarm
•
Logic: Selectable, positive or negative occurrence.
•
Indicator: Switchable horn and flashing display indicators.
Peak Detector
•
Amplitude: Peak amplitude of gated signal in percent of screen height.
•
Accuracy: ±1%.
Analog Output
•
Amplitude: Scalable. 1000 mV to 4000 mV with offset
-4000 mV to +4000 mV.
•
Accuracy: Greater of 30 mV or 2% of display indication with a 1000 ohm load.
Alarm Output
: 3.3 V logic signals (0V = no alarms, 3.3V = alarm condition).
Timebase:
Range
Range: 0.048 to 296" (1.23 mm to 7511mm) of steel (0.418 to 2560µs).
Accuracy: ±1%.
Delay
Range: -1.16 to 185" (-29 mm to 4690mm) of steel (-10 to 1600µs).
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...