Applications 95
Figures 5-28 and 5-29
4.
Press the Gate WIDTH
menu soft key
and rotate the SmartKnob™, or press the WIDTH
menu soft key
repeatedly, to adjust the Gate width until it goes past or through the 1st echo
from the thinnest step (see
Figure 5-29
)
.
The Gate width must extend past the thickest
step in the calibration process.
5.
Press the GATE2 menu key and the following menu will appear. This menu provides
changeable parameters, such as Polarity, Position, Width, and Level, for either flaw or
thickness measurements of Gate2.
For this application, Gate2 is used for thickness.
6.
To enable Gate2, press the GATE2 menu key again and rotate the SmartKnob™ to select
the “+” Gate2 alarm.
7.
Press the Gate LEVEL
menu soft key
and rotate the SmartKnob™ to the desired screen
height, such as 30%.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...