Technical Data 21
2. Technical Data
Pulser:
Type:
Square wave.
Pulse Width:
30 to 1000 ns (HR: 15 to 500 ns
Accuracy: ±10% or 10 ns, whichever is greater.
Pulse Voltage
: Selectable 300 or 150 volts.
Accuracy: +10/-15%.
Rise Time:
< 15 ns.
Fall Time:
< 60 ns damping 200 ohms. < 20 ns damping 50 ohms.
Damping
: Selectable 25, 50 or 200 ohms.
Modes:
Selectable single, dual, or through transmission.
Receiver:
Frequency
Band
: 0.3 to 20 MHz, +/-3 dB nominal.
Gain:
0 to100 dB in wideband, highpass, 10 MHz, and 15 MHz (HR). 0 to 110 dB in 1, 2.25,
and 5 MHz frequency settings.
Accuracy: ±0.5 dB for any 10 dB increment, cumulative errror < ±2.0 dB.
Control: Continuous adjustment in selectable 0.2, 1.0, 2.0, 6.0 or 12dB steps.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...