Applications 71
1.
Place couplant on the step block and position the transducer on the thinnest (0.100) step.
2.
Press RCVR menu key, then DISPLAY
menu soft key,
and rotate the SmartKnob™ to
select HALF-. Observe the signals and select “+” or “-” depending upon best signal
display.
3.
Adjust Delay by pressing the RANGE menu key, then DELAY
menu soft key
, and rotate
the SmartKnob™ to line up the leading edge of the first echo on the No. 1 graticule line.
4.
Adjust Gain by pressing the GAIN
menu soft key
, and rotate the SmartKnob™ to increase
the first echo amplitude to approximately 80% of full screen height (see
Figure 5-4
).
5.
Move the transducer to the thickest (0.500) step and press the RANGE menu key, then
select the VEL
menu soft key
, and rotate the SmartKnob™ to line up the leading edge of
the first echo on the #5 graticule line (see
Figure 5-5
).
6.
Repeat steps 1 - 5 to verify that both the 0.100 and 0.500 inch signals line up.
7.
The instrument should now be calibrated for 1.0 inch of steel across the full screen width.
Each major graticule division corresponds to 0.100 inch of steel.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...