114 Chapter 5
2.
Press the SELECT menu soft key; this will start the DAC Curve and log the point.
3.
Press the CURSOR
menu soft key
, and with the UP/DOWN arrow keys or SmartKnob™,
adjust the cursor to the second point/echo (see
Figure 5-53
).
4.
Press the SELECT
menu soft key
; this will log the second point/echo of the DAC Curve.
5.
Press the CURSOR
menu soft key
and with the UP/DOWN arrow keys or SmartKnob™,
adjust the cursor to the next point/echo.
6.
Press the SELECT
menu soft key
; this will log the next point/echo of the DAC Curve.
7.
Press the CURSOR
menu soft key
, and with the UP/DOWN arrow keys or SmartKnob™,
adjust the cursor to the last point/echo.
8.
Press the SELECT
menu soft key
; this will log the last point/echo of the DAC Curve (see
Figures 5-54 and 5-55
).The DAC Curve Only is now complete. If the curve is not
correct, repeat the above steps.
Follow the same steps above (DAC CURVONLY) for the 3 CURVE or 4 CURVE
mode of operation.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...