90 Chapter 5
Figure 22: First Back Echo at 0.100”
7.
Move the transducer to the thickest (0.500) step. Adjust the Velocity by pressing the
RANGE menu key, the VEL
menu soft key
, and rotating the SmartKnob™ to line up the
leading edge of the first echo on the last graticule line (refer to
Figure 5-23).
Figure 23: First Back Echo at 0.500”