Control Descriptions 37
Figure 4-1B: Sonic
®
1200HR Controls
4.2
Power Button
The POWER Button toggles instrument power on and off. Immediately after power is turned
ON, an initialization routine is started. The instrument’s microprocessor determines the
instrument configuration, performs power-up diagnostics and calibration, and readies the
instrument for operation. The instrument will normally attempt to restore its last configuration,
but can be made to load factory default settings by pressing and holding the RANGE key while
the instrument powers up.
4.3
Displays
The display contains a symbol indicating remaining battery life. When the unit has