Applications 89
To select and change the following parameters, press the
menu soft key
below the selected
menu box, then rotate the SmartKnob™ or press the INC/DEC arrow keys to change the value.
3.
Place couplant on the test block and position the transducer on the thinnest (0.100) step.
4.
Press the RCVR menu key, then the DISPLAY
menu soft key,
and rotate the SmartKnob™
to select HALF-. Observe the signals and select “+” or “-” depending upon best signal
display.
5.
Adjust Delay by pressing the RANGE menu key, then the DELAY
menu soft key
, and
rotate the SmartKnob™ to line up the leading edge of the first echo on the #2 graticule
line.
6.
Adjust Gain by pressing the RANGE menu key, then GAIN
menu soft key
, and rotate the
SmartKnob™ to increase the first echo amplitude to approximately 70% of full screen
height, #7 line vertical (refer to
Figure 5-22
).
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...