74 Chapter 5
8.
Place the transducer on the thinnest step (0.100), press the THK CAL softkey and rotate
the SmartKnob
™
until the THK CAL value for CAL PT2 reads 0.100 inch. Press the
ENTER key to log this value.
9.
Calibration is now done. Check the 0.200, 0.300, and 0.400 inch steps. If the thickness
readings are correct, the thickness gauge is calibrated. If the thickness gauge is not
calibrated, repeat calibration steps 5 through 8.
5.2
Dual Transducer Thickness Testing
Dual (pitch/catch) transducers produce better results for measuring thin (0.030 inch) materials
or rough and corroded metals in the 0.050 inch to 5.00 inch range. They can also be used on
smooth materials to more than 10 inches.
The following accessories are required for this setup procedure:
STEP BLOCK
— Sonic
®
#1920105.00 steel step block or equal (0.100 to 0.500 inch).
TRANSDUCER
— Harisonic
®
J4MD (¼ in. 5 MHz).
COUPLANT
— Use a couplant that “fills in” the surface. On smooth surfaces, use
thinner couplants (water, light oil, or mineral oil). On rough surfaces, the Sonic
®
X30S
couplant (P/N #3317450) is more appropriate.
5.2.1
Instrument Calibration - 0.100 to 0.500 Inches Steel
1.
Connect the two transducer leads into both BNC connectors.
2.
Turn instrument power on.
3.
Set the Sonic
®
1200S/HR for Dual mode by pressing the PULSER menu key and then
selecting the MODE
menu soft key.
4.
Adjust the operating controls as follows:
To select and change the following parameters, press the
menu soft key
below the selected
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...