Applications 97
12. Place the transducer on the 0.400 inch step. Verify proper Gate1 and Gate2 positioning, as
above (a slight adjustment of the Gate2 position may be required to prevent triggering on
unwanted signals), press the THK CAL softkey, and rotate the SmartKnob until the THK
CAL value for CAL PT2 reads 0.400 inch. Press ENTER to log this value.(Please refer to
Figure 5-33
).
13. Calibration is now done. Check other thickness values to verify; if the thickness readings
are correct, the thickness gauge is calibrated. If the thickness gauge is not calibrated,
repeat calibration steps 10 through 11.
Figure 5-32