Applications 75
menu box, then rotate the SmartKnob™ or press the INC/DEC arrow keys to change the value.
5.
Place X30S couplant on the step block and position the transducer on the thinnest (0.100)
step.
6.
Press the RCVR menu key, then the DISPLAY
menu soft key,
and rotate the SmartKnob™
to select HALF-. Observe the signals and select “+” or “-” depending upon which signal
displays the largest amplitude of the first cycle (see
Figure 5-8
)
.
7.
Adjust Delay by pressing the RANGE menu key, then the DELAY
menu soft key
, and
rotate the SmartKnob™ to line up the leading edge of the first echo on the #2 graticule
line.
8.
Adjust Gain by pressing the GAIN
menu soft key
, and rotate the SmartKnob™ to increase
the first echo amplitude to approximately 70% of full screen height (refer to
Figure 5-8
).
9.
Move the transducer to the thickest (0.500 inch) step. Adjust the Velocity by pressing the
RANGE menu key, the VEL
menu soft key
, and rotate the SmartKnob™ to line up the
leading edge of the first echo on the last graticule line (refer to
Figure 5-9).
10. Repeat steps 7 - 9 to verify that both the 0.100 and 0.500 inch signals line up.
11. Check the 0.200, 0.300. and 0.400 inch steps. If they line up, the Sonic
®
1200S/HR is
calibrated. If they do not line up closely, repeat calibration steps 5 through 9.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...