112 Chapter 5
Figure 5-50
g)
Press the TRIG
menu soft key
and turn ON the trig function by using the UP/DOWN
arrow keys or SmartKnob™.
h)
Press the ANGLE
menu soft key
and adjust the incidence angle to the desired value
using the UP/DOWN arrow keys, SmartKnob™, or by pressing the ANGLE menu
key repeatedly.
i)
Press the
THICK
menu soft key
and use either the UP/DOWN arrow keys, the
SmartKnob™, or press the THICK
menu
soft key
repeatedly to enter the thickness of
the part to be inspected. By entering the part thickness, the DEPTH of the reflector
will always be measured from the surface of the part, no matter what leg or V path
you are in.
j)
If the inspection is on a curved surface, press the O-DIAM
menu soft key
. Enter the
outer diameter of the curved surface by pressing the UP/DOWN arrow keys or
rotating the SmartKnob™. This corrects the trigometric functions for the curved
surface.
5.5
Distance Amplitude Correction (DAC) Testing
Press the DAC menu key, and then DAC
menu soft key,
and with the UP/DOWN arrow keys or
the SmartKnob, select one of the following modes of operation: 1) DAC CURVONLY; 2) 3
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...