Applications 73
3.
Press the Gate POSN
menu soft key
and rotate the SmartKnob™, or press the POSN
menu
soft key
repeatedly, to adjust the Gate position so that the start of the gate is to the left of
the thinnest (0.100 inch) step (see
Figure 5-6).
4.
Press the Gate WIDTH
menu soft key
and rotate the SmartKnob™, or press the WIDTH
menu soft key
repeatedly, to adjust the Gate width to full screen or to the right of the
thickest step (0.500 inch).
Figure 5-6: Echo From 0.100 Step and Figure 5-7: Echo From 0.500 Step
5.
Press the THICK menu key, then the T-GAUGE softkey, and the following menu will
display.
6.
Use the SmartKnob
™
to toggle through the selections. Select AIP - 1st. Press the
TRIGGER softkey and rotate the SmartKnob
™
to toggle between peak or leading edge
triggering for thickness measurements. Select EDGE.
7.
Place the transducer on the thickest step (0.500 inch of steel), press the THK CAL softkey,
and rotate the SmartKnob
™
until the THK CAL value for CAL PT1 reads 0.500 inch.
Press the ENTER key to log this value.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...