108 Chapter 5
5.4.3
Multiple Bounce Weld Inspection
If the sound path bounces or reflects one or more times from the sides of the test piece, the
sound path and surface distance readings will be the same as for direct reflection. However, the
depth calculation will give a result greater than the actual thickness (as shown in
Figure 5-45
).
Figure 5-45: Multiple Bounce Shear Calibration
1 = actual depth w = wall thickness
2 = calculated depth s = sound path (measured on EL/LCD
)
Depth calculation for one bounce: calculate the depth as in Direct Reflection.
depth (2) = S • cos (angle)
Subtract the depth (2) from twice the wall thickness to obtain the actual depth.
actual depth (1) = 2w - depth (2)
FLAW
WELD
W
Ø
1
2