Preparation for Operation 3
PowerLink™ Connection
After connecting a Olympus NDT transducer to your Sonic
®
1200S/HR instrument, touch and
hold
the transducer to the PowerLink contact disc on your EL/LCD display for approximately
4 seconds, or 2 beeps. If contact is not maintained for the proper amount of time, the screen
will show the PowerLink display with N/A parameters. In this case, try again, until the
PowerLink connection is complete and parameters are displayed.
Menu Description
•
Model
Transducer Model
•
Frequency
Transducer Nominal Frequency
•
S/N
Serial Number (stamped on the front of the transducer)
•
Mode
Single or Dual
•
Program Name and Date Stamp
- Program number, optional operator-entered description,
and date stamp. The stored program number is shown in the PROGRAM softkey menu
box.
Program
This line displays the program storage location/transducer setup, which is to be stored,
recalled, or erased. Rotating the SmartKnob
™
will cycle through the storage locations
associated with the identified transducer. If no stored locations are associated with the
identified transducer, the number of the next empty storage location will be displayed.
Store/Recall
Instrument settings associated with the identified transducer are stored to or recalled from the
selected program location. If the location is empty, the operator is allowed to store the
instrument setup. Whenever a transducer setup is recalled, the current instrument settings are
overwritten and cannot be recovered unless the setup was previously saved.
Erase
Allows the operator to erase the selected transducer setup. This needs to be done only if the
storage location is to be reused. An ENTER confirmation will be required from the operator
before the program location is cleared. An erased location cannot be recovered.
Name
Allows the operator to enter a name of up to 26 characters to identify the setup. Only program
locations which are not empty may be named.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...