76 Chapter 5
12. The Sonic
®
1200S/HR is now calibrated for steel from 0.100 inch to 0.500 inch This
method provides an accurate test over this range.
Figure 5-8. Pitch-Catch Echo from a 0.100 Step Block
Figure 5-9. Pitch-Catch Echo from a 0.500 Step Block
5.2.2
Instrument Calibration - 1.0 to 10.0 Inches of Steel
Use the same procedure as the previous 0.100 to 0.500 inch steel test, but adjust the controls
using 1 to 10 inch steel step blocks as the thin and thick steps. After calibrating at 1.0 inch and
10.0 inches, check at 5.0 inches for accuracy.
To select and change the following parameters, press the
menu soft key
below the selected
menu box, then rotate the SmartKnob™ or press the INC/DEC arrow keys to change the value.