Applications 99
5.4.1
Instrument Calibration - 10” Thickness Range, Steel
1.
Connect the transducer cable to the Spectrum
™
F2L transducer and to either BNC
connector.
2.
Turn on instrument power and adjust the operating controls as follows:
3.
Place couplant on the top surface of the test block at the index point matching the wedge
angle.
4.
Place transducer on a calibration block at the index point. Refer to
Figure 5-34
and
Figure 5-35
for IIW Blocks and
Figure 5-36
for the B1 block positions.
To select and change the following parameters, press the
menu soft key
below the
selected menu box, then rotate the SmartKnob™ or press the INC/DEC arrow keys to
change the value.