Control Descriptions 45
DELAY
Adjusts the time base delay from -1.16 to 185 inches in steel. Adjustment step size
varies from 0.001 to 1 unit.
Refer to the VEL description for further Delay limits.
VEL
Selects the material sound propagation velocity. The adjustment range is 0.0250 in/µs to
0.6000 in/µs. The allowable minimum and maximum values for range and delay are based on
the velocity setting.
If the instrument limits the adjustment for range, delay, gate position, or
gate width, it may be due to the current velocity setting
.
MAX REP
Adjusts the frequency of the square wave pulse applied to the transducer crystal.
The maximum rate is limited by the Range, Pulse Voltage, and Pulse Width settings.
Decreasing MAX REP may reduce the noise on the display.
4.14 Gate 1 Menu
The GATE 1 key calls up the gate 1 menu shown and is used to adjust the gate 1 parameters.
This menu sets up parameters such as polarity, position, width, and level for flaw gate 1. Gate
1 may also be used for thickness measurement by enabling the T-GAUGE control.
GAIN
See GAIN parameter description in the Receiver menu section
GATE 1
Selections are “+”, “-”, and “OFF”; “+” signals a gate alarm when the signal
amplitude is above the threshold level, and “-” signals an alarm when the flaw amplitude is
below the threshold level. “OFF” turns off the flaw gate. If a gate is in the alarm state when the
thickness gauge is OFF, the amplitude is displayed under the associated gate bar in inverse
video.
POSN
Adjusts the gate 1 start position from -1.160 to 185.0 inches in steel. The position is
displayed as the distance from IP in the selected units. Adjustment step size varies from 0.001
to 1 unit.
Refer to the VEL description for further Gate Position limits.
WIDTH
Adjusts the width of gate 1 from 0.001 to 296.0 inches in steel. Adjustment step size
is 0.001 to 1 unit.
Refer to the VEL description for further Gate Width limits.
LEVEL
Sets the gate 1 threshold value in terms of % of full screen height (FSH) for the alarm.
Adjustable from 5% to 100% of FSH in 1% increments.
GATE 1
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...