70 Chapter 5
3.
Place the transducer (with couplant) on the 1.0 inch thickness portion of the test block and
press the RANGE menu key.
4.
Press DELAY
menu soft key
and rotate SmartKnob™ to line up the 1st echo on #1
graticule line.
5.
Press VEL
menu soft key
and rotate SmartKnob™ to line up the 5th echo with #5 graticule
line.
6.
Repeat steps 3 - 5 until both echoes line up. All 10 echoes should then line up with a
graticule line, as shown in
Figure 5-3
. If they do, the instrument is calibrated for 1.0 to
10.0 inches. If not, repeat calibration.
Figure 5-3: Multiple Echoes From 1.0 Block
5.1.4
Digital Thickness Calibration for Contact Transducers
Adjust the operating controls as follows:
To select and change the following parameters, press the
menu soft key
below the
selected menu box, then rotate the SmartKnob™ or press the INC/DEC arrow keys to
change the value. These settings are approximate and may vary due to the different types of
transducers used.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...