Repeat steps 4 and 5 until both echoes line up. All 10 echoes should then line up with a
graticule line. If they do, the instrument is calibrated for 1.0 to 10.0 inches. If not, repeat
calibration.
7.
After calibration, you are ready to start testing for flaws. If the calibration block surfaces
are similar to the test material, you can proceed. If the surface (inner or outer) is rougher
or curved, you will need to adjust the Gain to acquire signals back up to about 70%.
Rough outer surfaces will also cause echoes after the initial pulse which restricts the
resolution of thin surfaces. Always make sure that the display signals are clear and “make
sense” before proceeding with the test.
5.2.4
Dual Element High/Low Digital Thickness Calibration
Adjust the operating controls as follows:
To select and change the following parameters, press the
menu soft key
below the selected
menu box, then rotate the SmartKnob™ or press the INC/DEC arrow keys to change the value.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...