Applications 109
Example
:
W = 1 inch and S = 3 inches and angle 0 = 60° then:
depth(2) = 3 • cos 60°
= 1.5”
actual depth(1) = (2 x 1”) - 1.5”
= 0.5”
Shear Wave (Angle Beam) Digital Thickness Calibration
Once the screen range is calibrated, continue with the setup of the Angle Beam (Trig functions)
mode. Adjust the operating controls as follows:
1.
1.For the correct thickness measurements, Gate1 must be adjusted as follows:
A IIW
Type 1 Block is used for this application.
2.
Press the GATE1 menu key, GATE1
menu soft key
, and press the UP/DOWN arrow keys
or rotate the SmartKnob™ until the “+ “ is selected.
3.
Press the LEVEL
menu soft key
and adjust the level to the desired screen height, such as
40%, with the UP/DOWN arrow keys or the SmartKnob™.
4.
Press the POSN
menu soft key
and adjust the position: the start of the gate is to the right of
the transducer wedge noise or IP, and to the left of the first echo (see
Figure 5-46
)
.
5.
Press the WIDTH
menu soft key
and adjust the width to full screen, or to the right of the
farthest echo, by rotating the SmartKnob
™
or pressing the UP/DOWN arrow keys (see
Figure 5-47
).
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...