Applications 85
4.
Press the Gate WIDTH
menu soft key
and rotate the SmartKnob™, or press the WIDTH
menu soft key
repeatedly, to adjust the Gate width to go past or through the 2nd back wall
echo (see
Figure 5-17
)
.
Figure 16 and 17
5.
For echo-to-echo measurements on thinner material than 0.500 inch, the Gate 2 position
(start) and width must be adjusted for the thinnest echo-to-echo measurement, such as
0.200 inch (see
Figure 5-18
).
6.
Gate 2 must start prior to the 2nd back wall echo (left side of echo) and the width must be
wide enough so the 2nd back wall echo passes through it (see
Figure 5-18
).
7.
Press the THICK menu key and the following menu will display. Press the T-GAUGE
menu soft key,
and use the UP/DOWN arrow keys to toggle through the selections. Select
Auto Echo - Echo.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...