72 Chapter 5
8.
Check the 0.200, 0.300 and 0.400 inch steps. If they line up, the Sonic
®
1200S/HR is
calibrated. If not, repeat calibration steps 1 - 5.
Figure 5-4: Echo From 0.100 Step and Figure 5-5: Echo From 0.500 Step
5.1.5
Digital Thickness Measurement From the Initial Pulse to
the 1st Back Wall Reflection
Press the GATE1 menu key and the following menu will appear. This menu provides
changeable parameters, such as Polarity, Position, Width, and Level, for either flaw or
thickness measurements of Gate1.
For thickness measurements, adjust the controls as follows:
1.
To enable Gate1, press GATE1 menu key again and rotate the SmartKnob™ to select the
“+” Gate1 alarm.
2.
Press the Gate LEVEL
menu soft key
and rotate the SmartKnob™ to the desired screen
height, such as 40%.
GATE 1
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...