58 Chapter 4
The A-Scan print function uses the optional external serial printer option to make a
hard copy of the stored A-Scan, program settings, and a custom report form.
ATTRIB
Attribute appears only when A scan location is ALL. Press ENTER to access the A
scan Attribute sub menu. when this soft key is highlighted.
4.25 A-Scan Attribute Sub Menu
The first character in the table above indicates the
attribute
of the A-Scan location.
•
[r]read-only
•
[e]extension of previous location
Setting the attribute of A-Scan locations protects these locations from accidentally being
erased. SET sets the attribute of the stored A-Scan, and CLEAR clears the attribute.
ATTRIB
This parameter displays the A-Scan memory location on which the attribute is to be
changed. Rotating the SmartKnob™ will cycle through the 100 A-Scan locations.
SET
The SET function is used to set the attribute of the highlighted A-Scan location. The A-
Scan can NOT be erased if the attribute is set.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...